Status structure
See Figure 4-5 for the Model 2000 Multimeters status structure. Instrument events, such as
errors, are monitored and manipulated by four status register sets. Notice that these status regis-
ter sets feed directly into the Status Byte Register. More detailed illustrations of these register
sets are provided by Figures 4-5 through 4-9.
0
2
3
5
6
Cal
7
9
10
11
12
13
14
15
Questionable
Condition
Register
(Always Zero)
0
2
3
5
6
7
9
10
11
12
13
14
15
Questionable
Event
Register
0
11 1
2
3
5
6
7
9
10
11
12
13
14
15
Questionable
Event
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
3
Trig
7
8
Idle
11
12
13
15
Operation
Condition
Register
(Always Zero)
3
7
8
Idle
11
12
13
15
Operation
Event
Register
3
7
8
Idle
11
12
13
15
Operation
Event
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
Idle
EAV
QSB
MAV
ESB
RQS/MSS
OSB
Status
Byte
Register
1
EAV
QSB
MAV
ESB
6
OSB
Service
Request
Enable
Register
&
&
&
&
&
&
&
Logical
OR
*STB?
*SRE
*SRE?
Master Summary Status (MSS)
MSB = Measurement Summary Bit
EAV = Error Available
QSB = Questionable Summary Bit
MAV = Message Available
ESB = Event Summary Bit
RQS/MSS = Request for Service/Master Summary Staus
OSB = Operation Summary Bit
Error Queue
Output Queue
Trig Trig
Note : RQS bit is in serial poll byte,
MSS bit is in *STB? response.
1
14 14
OPC
QYE
DDE
EXE
CME
URQ
PON
8
9
8
11
12
13
15
Standard
Event
Status
Register
8
9
8
11
12
13
15
Standard
Event
Status
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
(Always Zero)
Operation Complete
Query Error
Device Specific Error
Execution Error
Command Error
User Request
Power On
OPC
QYE
DDE
EXE
CME
URQ
PON
*ESR? *ESE
*ESE?
MSB MSB
12
13
14
15
(Always Zero)
12
13
14
15
Measurement
Event
Register
12
13
14
15
Measurement
Event
Enable
Register
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
&
Logical
OR
Measurement
Condition
Register
ROF
LL
HL
3
4
RAV
BAV
BHF
BFL
ROF
LL
HL
RAV
BAV
BHF
BFL
ROF
LL1
HL1
RAV
BAV
BHF
BFL
Reading Overfolw
Low Limit
High Limit
Reading Available
Buffer Available
Buffer Half Full
Buffer Full
11
Cal Cal
Calibration Summary
Command Warning
Warn Warn Warn
222
Temperature Summary
Temp Temp Temp
6
10
66
10 10
999
3
4
3
4
666
0
1
0
1
0
1
Meas Meas Meas
11 11 11
Measuring
Triggering
Remote Operation 4-19
Figure 4-5
Model 2000 status
register structure

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD