SCPI Command Reference 5-67
RS-232 interface commands
:LOCal
:SYSTem:LOCal Take 2000 out of remote
Description Normally, the Model 2000 is in local during RS-232 communications. In this
state, front panel keys are operational. However, the user may wish to lock
out front keys during RS-232 communications (see :RWLock).
This action command isu sed to take the Model 2000 out of the remote state
and enables the operation of front panel keys. Note that this command can
only be sent over the RS-232 interface.
:REMote
:SYSTem:REMote Place the Model 2000 in remote
Description This action command is used to place the Model 2000 in the remote state. In
remote, the front panel keys will be locked out if local lockout is asserted (see
:RWLock). Note that this command can only be sent over the RS-232 inter-
face.
:RWLock
:SYSTem:RWLock Disable front panel keys
Description This action command is used to disable front panel controls (local lockout)
during RS-232 operation.
Taking the instrument out of remote (see :LOCal) restores front panel keys
operation. Note that this command can only be sent over the RS-232 inter-
face.
Line frequency query
:LFRequency?
:SYSTem:LFRequency? Query line frequency
Description This query returns the frequency of the power line from which the unit is op-
erating. The power line frequency is automatically sensed upon power-up.

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
BrandKeithley
Model2000
CategoryMultimeter
LanguageEnglish

Related product manuals