Connections
Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the INPUT HI and LO terminals of the Model 2000. Either the
front or rear inputs can be used; place the INPUTS button in the appropriate position.
2. Select the FREQ or PERIOD function.
3. Connect test leads to the source as shown in Figure 2-7.
CAUTION Do not exceed 1000V peak between INPUT HI and INPUT LO or instru-
ment damage may occur.
4. Take a reading from the display.
See Section 3—Measurement Options for information that explains the configuration options
for frequency and period measurements.
2001 MULTIMETER
Model 2000
Caution: Maximum Input = 1000V peak, 8 x 10
7
V•Hz
AC Voltage
Source
Input Impedance = 1MΩ in parallel with <100pF
SHIFT
CH1REM
TALK
LSTN
SRQ
STATREL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH
8
CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Basic Measurements 2-27
Figure 2-7
Frequency and
period measure-
ments