EasyManua.ls Logo

Keithley 2000 - Two- and Four-Wire Resistance Measurements

Keithley 2000
280 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Shielding
To achieve a stable reading, it helps to shield resistances greater than 100k. Place the resis-
tance in a shielded enclosure and connect the shield to the INPUT LO terminal of the instrument
electrically.
See Section 3—Measurement Options for information that explains the configuration options
for 2-wire and 4-wire resistance measurements.
2001 MULTIMETER
Model 2000
Resistance
Under Test
Shielded
Cable
Optional shield
Note: Source current flows from the INPUT
HI to INPUT LO terminals.
SHIFT
CH1REM
TALK
LSTN
SRQ
STATREL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH
8
CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
2001 MULTIMETER
Resistance
Under Test
Shielded
Cable
Optional shield
Note: Source current flows from the INPUT
HI to INPUT LO terminals.
Model 2000
SHIFT
CH1REM
TALK
LSTN
SRQ
STATREL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH
8
CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
Basic Measurements 2-25
Figure 2-6
Two- and four-
wire resistance
measurements

Table of Contents

Other manuals for Keithley 2000

Related product manuals