Taking readings with the scanner card
The Model 2000-SCAN is an optional 10-channel scanner card for the Model 2000 Multim-
eter. Only one channel can be closed at a time. If you close a channel while another is already
closed, the first one opens with break-before-make operation.
You can use the scanner card two ways. One is to issue a command to close a particular chan-
nel before sending other commands to take readings. The other way is to program the scan list,
and let the meter take care of closing a channel before taking a reading.
The following example program measures DC volts on channel 1, AC volts on channel 2, and
2-wire resistance on channel 3, using the
ROUTe:CLOSe
command.
C-8 Example Programs

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)