EasyManua.ls Logo

Keithley 2000 - Speed Commands

Keithley 2000
280 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Speed Commands
:NPLCycles <n>
[:SENSe[1]]:CURRent:AC:NPLCycles <n> Set NPLC for ACI
[:SENSe[1]]:CURRen[:DC]:NPLCycles <n> Set NPLC for DCI
[:SENSe[1]]:VOLTage:AC:NPLCycles <n> Set NPLC for ACV
[:SENSe[1]]:VOLTage[:DC]:NPLCycles <n> Set NPLC for DCV
[:SENSe[1]]:RESistance:NPLCycles <n> Set NPLC for 2
[:SENSe[1]]:FRESistance:NPLCycles <n> Set NPLC for 4
[:SENSe[1]]:TEMPerature:NPLCycles <n> Set NPLC for TEMP
Parameters <n> = 0.01 to 10 Power line cycles per integration
DEFault 1
MINimum 0.01
MAXimum 10
Query :NPLCycles? Query programmed NPLC value
:NPLCycles? DEFault Query *RST default value
:NPLCycles? MINimum Query minimum NPLC value
:NPLCycles? MAXimum Query maximum NPLC value
Description The integration period (measurement speed) for the basic measuement func-
tions (except Frequency and Period) is set using the :NPLCycle command.
NPLC (Number of Power Line Cycles) expresses the integration period by
basing it on the power line frequency. For example, for a PLC of 1, the inte-
gration period in seconds would be 1/60 (for 60Hz line power) which is
16.67msec.
:APERture <n>
[:SENSe[1]]:FREQuency:APERture <n> Set aperture for frequency
[:SENSe[1]]:PERiod:APERture <n> Set aperture for period
Parameter <n> = 0.01s to 1.0s
DEFault 1.0s
MINimum 0.01s
MAXimum 1.0s
Query :APERture? Query programmed APERture value
Description The integration period (measurement speed) for Frequency and Period is set us-
ing the APERture command (to set other basic measurement functions, see
:NPLCycle). The unit of measurement for the APERture command is sec-
onds.