Measuring current
The Model 2000 can make DCI measurements from 10nA to 3A and ACI measurements from
1µAm to 3A RMS.
NOTE See the previous discussion about crest factor in “Measuring voltage” in this section.
Connections
Assuming factory default conditions, the basic procedure is as follows:
1. Connect test leads to the AMPS and INPUT LO terminals. The front inputs must be
used; place the INPUTS button in the FRONT position.
2. Select the measurement function by pressing DCI or ACI.
3. Pressing AUTO toggles autoranging. Notice the AUTO annunciator is displayed with
autoranging. If you want manual ranging, use the RANGE and keys to select a
measurement range consistent with the expected current.
4. Connect test leads to the source as shown in Figure 2-5.
CAUTION Do not apply more than 3A, 250V to the input or the AMPS fuse will open-
circuit.
5. Observe the display. If the “OVERFLOW” message is displayed, select a higher range
until a normal reading is displayed (or press AUTO for autoranging). Use the lowest pos-
sible range for the best resolution.
6. Take readings from the display.
2001 MULTIMETER
Model 2000
Caution: Maximum Input = 3A DC or RMS
Current
Source
SHIFT
CH1REM
TALK
LSTN
SRQ
STATREL FILT
4W
BUFFER
MATH
REAR
SCAN
TIMER
STEP CH2 CH3 CH4 CH5 CH6 CH7 CH
8
CH9 CH1
0
HOLD TRIG FAST MED SLOW AUTO ERR
2-22 Basic Measurements
Figure 2-5
DC and AC cur-
rent measure-
ments

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
BrandKeithley
Model2000
CategoryMultimeter
LanguageEnglish

Related product manuals