Program examples
All examples presume QuickBASIC version 4.5 or higher and a CEC IEEE-488 interface
card with CEC driver version 2.11 or higher, with the Model 2000 at address 16 on the IEEE-
488 bus.
Changing function and range
The Model 2000 has independent controls for each of its measurement functions. This means,
for example, that autorange can be turned on for DC voltage while leaving it off for AC voltage.
Another difference is in the parameter to the range command. In other instruments, a single
number was used to denote each range. The parameter of the SCPI
RANGe
command is given as
"the maximum value to measure." The instrument interprets this parameter and goes to the ap-
propriate range. When you query the range with
RANGe?
the instrument sends back the full-scale
value of its present range.
The following example program illustrates changing function and range. It sets the range for
several functions, then takes readings on each of those functions.
Note that the Model 2000 rounds the range parameter to an integer before choosing the ap-
propriate range. Sending
VOLTage:DC:RANGe 20.45
will set the Model 2000 to the 100V
range.
C-2 Example Programs

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD

Related product manuals