:REAL:OFFSet <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:REAL:OFFSET <n>
Parameters <n> = -0.09999 to 0.09999 Specify voltage offset at 0°C
DEFault 0.05463
MINimum -0.09999
MAXimum 0.09999
Query :OFFSet? Query voltage offset
:OFFSet? DEFault Query *RST default voltage offset
:OFFSet? MINimum Query lowest allowable voltage offset
:OFFSet? MAXimum Query largest allowable voltage offset
Description This command is used to specify the offset voltage at 0°C for the specified
reference junction.
:DIODe command
:RANGe[:UPPer] <NRf>
[:SENSe[1]]:DIODe:CURRent:RANGe[:UPPer] <NRf> Select current range for diode test
Parameters <NRf> = 0 to 1e-3 Specify diode test current
Query [UPPer]? Query selected range
Description There are three current ranges available for the diode test: 10µA range,
100µA range and the 1mA range. Range is selected by using this command
to specify the expected current for the diode under test. The instrument will
then automatically select the appropriate range.
:CONTinuity command
:THReshold <n>
[SENSe[1]]:CONTinuity:THReshold <NRf> Specify threshold resistance
Parameters <NRf> = 1 to 1000 Specify threshold in ohms
Query :THReshold? Query threshold resistance
This command is used to specify the threshold resistance for the continuity
test. Continuity occurs when the measurement is less than or equal to the
specified threshold level.