:SIMulated <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:SIMulated <n>
Parameters <n> = 0 to 50 Specify temperature in °C
32 to 122 Specify temperature in °F
273 to 323 Specify temperture in K
DEFault 23°C, 73.4°F, 296K
MINimum 0°C, 32°F, 273K
MAXimum 50°C, 122°F, 323K
Query :SIMulated? Query simulated reference
:SIMulated? DEFault Query default *RST reference
:SIMulated? MINimum Query lowest allowable reference
:SIMulated? MAXimum Query largest allowable reference
Description This command is used to specify the simulated reference temperature. The
temperature value depends on which temperature scale is currently selected
(°C, °F or K). Typically, 0° or 23°C is used as the simulated reference tem-
perature.
:REAL:TCOefficient <n>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]]:REAL:TCoefficient <n>
Parameters <n> = -0.09999 to 0.09999 Specify temperature coefficient
DEFault +0.01 temperature coefficient
MINimum -0.09999 temperature coefficient
MAXimum +0.09999 temperature coefficient
Query :TCOefficient? Query temperature coefficient (TC)
:TCOefficient? DEFault Query *RST default TC
:TCOefficient? MINimum Query lowest allowable TC
:TCOefficient? MAXimum Query largest allowable TC
Description This command is used to specify the temperature coefficient (TC) of the “re-
al” temperature reference junction. TC is specified in °C/volt and is not af-
fected by the :UNIT:TEMPerature command.
5-50 SCPI Command Reference