SCPI Command Reference 5-57
RAV
B15 - B12 B10
B9 B8 B7 B6 B5
B4 B3
B2 B1
B0
(2 )
9
(2 )
8
(2 )
7
(2 )
5
(2 )
2
(2 )
1
(2 )
0
32
0/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Measurement Event
0 = Disable (Mask) Measurement Event
LL
2
0/1
Events : BFL = Buffer Full
BHF = Buffer Half Full
BAV = Buffer Available
RAV = Reading Available
HL = High Limit
LL = Low Limit
ROF = Reading Overflow
BFL HL ROF
512 4 1
0/1 0/1 0/1
BHF BAV
256 128
0/1 0/1
B11
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Questionable Event
0 = Disable (Mask) Questionable Event
Events : Warn = Command Warning
Cal = Calibration Summary
Temp = Temperature Summary
Cal
B13 - B9 B8
B15 B14
(2 )
8
256
(2 )
14
16384
0/10/1
Warn
B7 - B5
0
Temp
B4
(2 )
4
16
0/1
B3 - B0
Idle MeasTrig
B15 - B11 B10 B9 B8 B7 B6 B5 B4 B3 B2 B1
B0
(2 )
10
1024
(2 )
4
(2 )
5
1632
0/1 0/10/1
Bit Position
Event
Decimal Weighting
Value
Value : 1 = Enable Operation Event
0 = Disable (Mask) Operation Event
Events : Idle = Idle state of the 2000
Trig = Triggering
Meas = Measuring
Figure 5-7
Measurement
event enable reg-
ister
Figure 5-8
Questionable
event enable reg-
ister
Figure 5-9
Operation event
enable register