Table 5-11
UNIT command summary
Command Description
Default
parameter
SCPI
:UNIT
:TEMPerature <name>
:TEMPerature?
:VOLTage
:AC <name>
:DB
:REFerence <n>
:REFerence?
:DBM
:IMPedance <n>
:IMPedance?
:AC?
[:DC] <name>
:DB
:REFerence <n>
:REFerence?
:DBM
:IMPedance <n>
:IMPedance?
:DC?
Select temperature measurement units (C, F, or K).
Query temperature units.
Path to configure voltage units.
Select ACV measurement units (V, DB or DBM).
Path to set DB reference voltage.
Specify reference in volts (le-7 to 1000).
Query DB reference.
Path to set DBM reference impedance.
Specify reference impedance (1 to 9999).
Query DBM reference impedance.
Query ACV units.
Select DCV measurement units (V, DB, or DBM)
Path to set DB reference voltage:
Specify reference in volts (1e-7 to 1000).
Query reference.
Path to set DBM refernece impedance:
Specify reference impedance (1 to 9999).
Query reference impedance.
Query DCV units.
C
V
1
75
V
1
75
SCPI Command Reference 5-19