Thermocouple commands
:TYPE <name>
[:SENSe[1]]:TEMPerature:TCouple:TYPE <name> Specify TC type
Parameters <name> = J Set operation for Type J thermocouples
K Set operation for Type K thermocouples
T Set operation for Type T thermocouples
Query :TYPE? Query thermocouple type
Description This command is used to configure the Model 2000 for the thermocouple
type that you are using to make temperature measurements.
These commands are used to configure the reference junction for thermocou-
ple temperature measurements.
:RSELect <name>
[:SENSe[1]]:TEMPerature:TCouple:RJUNction[1]:RSELect <name> Specify reference junction type.
Parameters <name> = SIMulated Use simulated temperature as reference
REAL Use a measured temperature as reference
Query :RSELect? Query reference junction type
Description This command is used to specify the type of reference junction that is going
to be used for thermocouple temperature measurements. Specify REAL if
you are using an actual reference junction. The :REAL command is then
used to specify the desired reference temperature. Specify SIMulated if you
wish to use a simulated reference temperature. The :SIMulated command is
then used to specify the desired simulated reference temperature.
SCPI Command Reference 5-49