5-58 SCPI Command Reference
:CONDition? command
:CONDition?
:STATus:MEASurement:CONDition? Read Measurement Condition Register
:STATus:QUEStionable:CONDition? Read Questionable Condition Register
:STATus:OPERation:CONDition? Read Operation Condition Register
Description
These query commands are used to read the contents of the condition regis-
ters. Each set of event registers (except the Standard Event register set) has a
condition register. A condition register is similar to its corresponding event
register, except that it is a real-time register that constantly updates to reflect
the current operating status of the instrument. See [:EVENt] for register bit
descriptions.
After sending one of these commands and addressing the Model 2000 to talk,
a decimal value is sent to the computer. The binary equivalent of this decimal
value indicates which bits in the register are set.
For example, if sending :stat:meas:cond? returns a decimal value of 512 (bi-
nary 0000001000000000), bit B9 of the Measurement Condition Register is
set indicating that the trace buffer is full.
:PRESET command
:PRESET
:STATUS:PRESET Return registers to default conditions
Description
When this command is sent, the SCPI event registers are affected as follows:
All bits of the following registers are cleared to zero (0):
Questionable Event Enable Register.
Measurement Event Enable Register.
Operation Event Enable Register
NOTE
Registers not included in the above list are not affected by this
command.

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
Digits6.5
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
Resistance Range100 Ω to 100 MΩ
ConnectivityIEEE-488 (GPIB)
Display TypeVFD