:AUTO <b>
[:SENSe[1]]:CURRent:AC:RANGe:AUTO <b> Control auto range for ACI
[:SENSe[1]]:CURRent[:DC]:RANGe:AUTO <b> Control auto range for DCI
[:SENSe[1]]:VOLTage:AC:RANGe:AUTO <b> Control auto range for ACV
[:SENSe[1]]:VOLTage[:DC]:RANGe:AUTO <b> Control auto range for DCV
[:SENSe[1]]:RESistance:RANGe:AUTO <b> Control auto range for 2
[:SENSe[1]]:FRESistance:RANGe:AUTO <b> Control auto range for 4
Parameters <b> = 1 or ON Enable auto range
0 or OFF Disable auto range
Query :AUTO? Query auto range (on or off)
Description These commands are used to control auto ranging. With auto ranging en-
abled, the instrument automatically goes to the most sensitive range to per-
form the measurement.
The auto range command (:RANGe:AUTO) is coupled to the command that
manually selects the measurement range (:RANGe <n>). When auto range
is enabled, the parameter value for :RANGe <n> changes to the automati-
cally selected range value. Thus, when auto range is disabled, the instrument
remains at the automatically selected range. When a valid :RANGe <n>
command is sent, auto ranging disables.