:COUNt <n>
[:SENSe[1]]:CURRent:AC:AVERage:COUNt <n> Specify filter count for ACI
[:SENSe[1]]:CURRent[:DC]:AVERage:COUNt <n> Specify filter for DCI
[:SENSe[1]]:VOLTage:AC:AVERage:COUNt <n> Specify filter count for ACV
[:SENSe[1]]:VOLTage[:DC]:AVERage:COUNt <n> Specify filter count for DCV
[:SENSe[1]]:RESistance:AVERage:COUNt <n> Specify filter count for 2
[:SENSe[1]]:FRESistance:AVERage:COUNt <n> Specify filter count for 4
[:SENSe[1]]:TEMPerature:AVERage:COUNt <n> Specify filter count for TEMP
Parameters <n> = 1 to 100 Specify filter count
DEFault 10
MINimum 1
MAXimum 100
Query :COUNt? Query filter count
:COUNt? DEFault Query the *RST default filter count
:COUNt? MINimum Query the lowest allowable filter count
:COUNt? MAXimum Query the larges allowable filter count
Description These commands are used to specify the filter count. In general, the filter
count is the number of readings that are acquired and stored in the filter buff-
er for the averaging calculation. The larger the filter count, the more filtering
that is performed.
SCPI Command Reference 5-47

Table of Contents

Other manuals for Keithley 2000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2000 and is the answer not in the manual?

Keithley 2000 Specifications

General IconGeneral
BrandKeithley
Model2000
CategoryMultimeter
LanguageEnglish

Related product manuals