Keysight M8070A Programming Guide 115
SCPI Command Reference 5
:STATus:QUEStionable:PTRansition[?]
Syntax :STATus:QUEStionable:PTRansition <NRf>
:STATus:QUEStionable:PTRansition?
Input
Parameters
<NRf>: Set transition filter state.
Description This command sets the transition filter state in the Questionable Status
Register structure. This is the default setting of the instrument. The query
returns the weighted value of the bits that are set to pass positive
transitions in the transition filter.
:STATus:QUEStionable:INSTrument
The summary bit of this register set is reflected in bit 9 of the Questionable
status register.
The :STATus:QUEStionable:INSTrument contains conditions which reflect
states of the instrument’s normal operation. These bits are summary bits.
For example, if a Data Loss is detected by one or more error detectors, the
Data Loss condition will be set.
The definition of each of these bits (condition register) is as follows:
Bit 0 – Error
One or more error detectors detect an error.
Bit 1 – Data Loss
This bit is set when the data source is turned off, not connected, or the
cables or device is faulty. This bit can also be set when the 0/1 threshold is
not within the eye limits of the incoming data signal. In this last case, use
Auto Align or select Avg 0/1 Threshold.
Bit 2 – Sync Loss
This bit is set when the error detector pattern does not match the
incoming data pattern or the BER of your device is higher than the sync
threshold.
Bit 3 – Clock Loss
This bit is set when the pattern generator receives no external clock signal
or the error detector receives no clock input signal.
Bit 4 – CDR Unlocked
One or more CDRs of the instrument are unlocked.

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals