Keysight M8070A Programming Guide 131
SCPI Command Reference 5
:TRIGger[:SOURce]:FREQuency?
Syntax :TRIGger[:SOURce]:FREQuency? ‘identifier’
Input
Parameters
‘identifier’: ‘M1.TrigOut’
Description This query returns the effective frequency of the trigger out clock signal.
This parameter is only active if the subrate clock operating mode is
selected. The mentioned frequency value depends upon the system
frequency & divider value.
This SCPI is applicable for M8041A.
Example :TRIG:FREQ? ‘M1.TrigOut’
:TRIGger:INTernal[:SOURce][?]
Syntax :TRIGger:INTernal[:SOURce] ‘identifier’, <AXIFrame|INTernal>
:TRIGger:INTernal[:SOURce]? ‘identifier’
Input
Parameters
‘identifier’: M1.ClkGen
<AXIFrame>: Select 100 MHz reference of AXIFrame as reference clock.
<INTernal>: Select 100 MHz reference of the M8041A as reference clock.
Return Range AXIF|INT
Description In the INTernal mode different reference clock sources are available for the
internal oscillator. Select between the 100 MHz reference clock of the
M9505A AXIe Chassis or the 100 MHz reference clock of the M8041A
module.
This SCPI is applicable for M8041A and M8045A.
Example :TRIGger:INTernal:SOURce ‘M1.ClkGen’,INT
:TRIGger:INTernal:SOURce? ‘M1.ClkGen’

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals