130 Keysight M8070A Programming Guide
5 SCPI Command Reference
:TRIGger[:SOURce][?]
Syntax :TRIGger[:SOURce] ‘identifier’, INTernal|REFerence|DIRect|CMULtiplier
:TRIGger[:SOURce]? ‘identifier’
Input
Parameters
‘identifier’: M1.ClkGen
Return Range INT|REF|DIR|CMUL
Description Selects the different trigger modes. The query returns the current trigger
source mode.
The trigger source modes include the following:
INTernal 100 MHz reference clock or AXIe 100 MHz (M9505A AXIe
Chassis)
REFerence 10 MHz / 100 MHz external reference clock
DIRect 8.1 GHz to 16.207 GHz clock used as the system frequency
directly in the M8041A
CMULtiplier Multiplied/divided clock used to specify system frequency
This SCPI is applicable for M8041A, M8062A and M8045A.
Example :TRIGger:SOURce ‘M1.ClkGen’, INT
:TRIGger:SOURce? ‘M1.ClkGen’
:CMULtiplier:FREQuency[?] :TRIGger:CMULtiplier:FREQuency:DETect on
page 135
:CMULtiplier:FREQuency:DETect :TRIGger:CMULtiplier:FREQuency:DETect on
page 135
:CMULtiplier:FREQuency:MULTipli
er[?]
:TRIGger:CMULtiplier:FREQuency:MULTipli
er[?] on page 133
:CMULtiplier:FREQuency:DIVider[
?]
:TRIGger:CMULtiplier:FREQuency:DIVider[?
] on page 134
:CMULtiplier:FREQuency[?] :TRIGger:CMULtiplier:FREQuency[?] on
page 134
:CMULtiplier:FREQuency:DETect :TRIGger:CMULtiplier:FREQuency:DETect on
page 135
Name Description under