Keysight M8070A Programming Guide 143
SCPI Command Reference 5
Description This command sets the time from the start of the period to the first edge of
the pulse. Acceptable units include ps, ns and exponents (for example,
25E-9 is the same as 25 ns).
This SCPI is applicable for M8041A, M8051A, M8195A, M8062A and
M8045A.
Example :PULS:DEL ‘M1.DataOut2’, 25ns
[:SOURce]:PULSe:TRANsition:FIXed[?]
Syntax [:SOURce]:PULSe:TRANsition:FIXed ‘identifier’, <SMOoth | MODerate |
STEep>
[:SOURce]:PULSe:TRANsition:FIXed? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’, ‘M*.DataOut2’, or ’M*.DataOut’
<SMOoth | MODerate | STEep>
Return Range SMOoth | MODerate | STEep
Description Use this command to set/read the transition time of the data output
stream.
Following is the expected transition time (tr) for the below
mentioned settings:
1. Smooth: 20 ps typical (20%-80%)
2. Moderate: 17 ps typical (20%-80%)
3. Steep: 12 ps typical (20%-80%)
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PULS:TRAN:FIX 'M1.DataOut2', SMO
:PULS:TRAN:FIX? 'M1.DataOut2'
[:SOURce]:PULSe:TRANsition[:LEADing][?]
Syntax [:SOURce]:PULSe:TRANsition[:LEADing] ‘identifier’,<NRf>
[:SOURce]:PULSe:TRANsition[:LEADing]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1’,‘M*.DataOut2’, ‘M*.DataOut3’ and ‘M*.DataOut4’
<NRf>: Specify transition time value.