Keysight M8070A Programming Guide 161
SCPI Command Reference 5
[:SOURce]:JITTer:HFRequency:BUNCorrelate:PRBSequence[:SELect][?]
Syntax [:SOURce]:JITTer:HFRequency:BUNCorrelate:PRBSequence[:SELect]
‘identifier’,
<PRBS7|PRBS8|PRBS9|PRBS10|PRBS11|PRBS15|PRBS23|PRBS31
[:SOURce]:JITTer:HFRequency:BUNCorrelate:PRBSequence[:SELect]?
‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or 'M*.DataOut’
<PRBS7|PRBS8|PRBS9|PRBS10|PRBS11|PRBS15|PRBS23|PRBS31>:
Select the PRBS polynomial.
Return Range PRBS7|PRBS8|PRBS9|PRBS10|PRBS11|PRBS15|PRBS23|PRBS31
Description This command selects polynomial of the PRBS for bounded uncorrelated
jitter source.
This query returns the current setting.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :JITT:HFR:BUNC:PRBS ‘M2.DataOut2’, PRBS7