170 Keysight M8070A Programming Guide
5 SCPI Command Reference
[:SOURce]:JITTer:LFRequency:UNIT[?]
Syntax [:SOURce]:JITTer:LFRequency:UNIT ‘identifier’, <UINTerval|TIME>
[:SOURce]:JITTer:LFRequency:UNIT? ‘identifier’
Input
Parameters
‘identifier’: 'M1.ClkOut', 'M*.DataOut1', 'M*.DataOut2', or 'M*.DataOut'
<UINTerval|TIME>: Specify jitter parameter units.
Return Range UINT|TIME
Description Specifies whether the jitter parameters are to be specified/returned in
seconds (TIME) or unit intervals (UINTerval).
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :JITT:LFR:UNIT ‘M1.DataOut2’,TIME