Page #200 background image
200 Keysight M8070A Programming Guide
5 SCPI Command Reference
[:SOURce]:INTerference:RANDom:CREStfactor:UNIT[?]
Syntax [:SOURce]:INTerference:RANDom:CREStfactor:UNIT[?] ‘identifier’,
< LINear | LOGarithmic >
[:SOURce]:INTerference:RANDom:CREStfactor:UNIT? ‘identifier’
Input
Parameters
‘identifier’: ‘M*.DATAOUT1’ or ‘M*.DataOut2’
< LINear | LOGarithmic >
Description This command sets the random interference crest factor unit type.
Following unit types are allowed:
• LINear - This is the default setting. Select this unit to specify the crest
factor as voltage ratio.
• LOGarithmic - Select this unit to specify the crest factor in dB.
This SCPI is applicable for M8195A and M8196A.
Example :INT:RAND:CRES:UNIT ‘M2.DataOut1’, LOG