Keysight M8070A Programming Guide 205
SCPI Command Reference 5
:OUTPut:GLOBal[:STATe][?]
Syntax :OUTPut:GLOBal[:STATe] ‘identifier’, <ON|OFF|1|0>
:OUTPut:GLOBal][:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M1.System'
<ON|OFF|1|0>: Enable/disable global state.
Return Range 1|0
Description This command sets global output state to ON or OFF.
This command works for all modules, as it is a Global command.
Example :OUTP:GLOB ‘M1.System’, off
:OUTPut[:STATe][?]
Syntax :OUTPut[:STATe] ‘identifier’, <ON|OFF|1|0>
:OUTPut[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M1.ClkOut', 'M1.TrigOut', 'M1.SysOutA' or 'M1.SysOutB',
'M*.DataOut1', 'M*.DataOut2', 'M*.CtrlOut', ‘M*.CtrlOutA’, or ‘M*.DataOut’
<ON|OFF|1|0>: Switch an output on or off.
Return Range 1|0
Description This command switches an output on or off.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8195A
and M8045A.
Example :OUTP ‘M1.DataOut1’, off
:OUTPut:CONFig:SOURce[?]
Syntax :OUTPut:CONFig:SOURce <Identifier>, <CLEanclock|DATaclock>
:OUTPut:CONFig:SOURce? <Identifier>
Input
Parameters
‘identifier’: ‘M1.ClkOut1’ or ‘M1.ClkOut2’
Return Range CLE|DAT