206 Keysight M8070A Programming Guide
5 SCPI Command Reference
Description This command selects the signal source for the instrument’s output. This is
only valid for certain clock outputs.
This query returns the present state.
This SCPI is supported by M8045A.
Example :OUTP:CONF:SOUR ‘M1.ClkOut1’,CLE
:OUTPut:DRATe?
Syntax :OUTPut:DRATe? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', 'M*.DataOut', or ‘M1.ClkOut’
Description This query returns data rate for the output channel.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8195A
and M8045A.
Example :OUTP:DRAT? 'M1.DataOut1'
:OUTPut:POLarity[?]
Syntax :OUTPut:POLarity ‘identifier’, <NORMal|INVerted>
:OUTPut:POLarity? ‘identifier’
Input
Parameters
‘identifier’: 'M1.TrigOut', 'M*.DataOut1', 'M*.DataOut2', 'M*.CtrlOutA',
'M*.SysOutA', 'M*.SysOutB', or 'M*.DataOut'
<NORmal|INVerted>: Set output polarity to normal or inverted.
Return Range NORM|INV
Description This command sets the output polarity to either normal or inverted.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :OUTP:POL ‘M1.DataOut1’, INV
:OUTPut:TVOLtage[?]
Syntax :OUTPut:TVOLtage ‘identifier’, <NRf>
:OUTPut:TVOLtage? ‘identifier’