EasyManuals Logo

Keysight Technologies J-BERT M8020A User Manual

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #206 background imageLoading...
Page #206 background image
206 Keysight M8070A Programming Guide
5 SCPI Command Reference
Description This command selects the signal source for the instrument’s output. This is
only valid for certain clock outputs.
This query returns the present state.
This SCPI is supported by M8045A.
Example :OUTP:CONF:SOUR ‘M1.ClkOut1’,CLE
:OUTPut:DRATe?
Syntax :OUTPut:DRATe? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', 'M*.DataOut', or ‘M1.ClkOut’
Description This query returns data rate for the output channel.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8195A
and M8045A.
Example :OUTP:DRAT? 'M1.DataOut1'
:OUTPut:POLarity[?]
Syntax :OUTPut:POLarity ‘identifier’, <NORMal|INVerted>
:OUTPut:POLarity? ‘identifier’
Input
Parameters
‘identifier’: 'M1.TrigOut', 'M*.DataOut1', 'M*.DataOut2', 'M*.CtrlOutA',
'M*.SysOutA', 'M*.SysOutB', or 'M*.DataOut'
<NORmal|INVerted>: Set output polarity to normal or inverted.
Return Range NORM|INV
Description This command sets the output polarity to either normal or inverted.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :OUTP:POL ‘M1.DataOut1’, INV
:OUTPut:TVOLtage[?]
Syntax :OUTPut:TVOLtage ‘identifier’, <NRf>
:OUTPut:TVOLtage? ‘identifier’

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals