210 Keysight M8070A Programming Guide
5 SCPI Command Reference
Return Range OFF|IDLE|EXT
Description This command enables/disables the electrical idle feature for the Data
Output. The state can be OFF, IDLE or EXTernal.
This SCPI is applicable for M8061A and M8062A.
Example :OUTP:EIDL ‘M2.DataOut’, EXT
:OUTPut:DATA:XOVer[?]
Syntax :OUTPut:DATA:XOVer ‘identifier’, <NRf>
:OUTPut:DATA:XOVer? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or 'M*.DataOut'
<NRf>: Specify an eye crossover.
Return Range For M8041A and M8051A: 10% to 90%
For M8061A and M8062A: 30% to 70%
Description This command sets the eye crossover of the transmitter signal at the DATA
OUT/-DATA OUT connectors. Crossover is defined for NRZ signals.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example :OUTP:DATA:XOV ‘M1.DataOut2’, 25
:OUTPut:DATA:F2Jitter[?]
Syntax :OUTPut:DATA:F2Jitter ‘identifier’, <NRf>
:OUTPut:DATA:F2Jitter? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or 'M*.DataOut'
<NRf>: Set the f/2 jitter.
Return Range -20 ps to +20 ps
Description This command sets the f/2 jitter at the output in seconds. Acceptable units
include ps and exponents (for example, -10e-12 is the same as -10 ps).
This query returns the present setting.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8045A.
Example :OUTP:DATA:F2J ‘M1.DataOut2’, -10e-12

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Summary

Keysight M8000 Series of BER Test Solutions

Safety Summary

Covers general safety precautions, environment conditions, and handling instructions for instrument operation.

1 Programming Basics

Modular Configuration

Discusses the modular configuration of instruments for remote programming.

Using Identifiers

Explains the use of Location, Group Name, and Measurement Name identifiers.

2 Recommended Programming Techniques

Output Protection

Covers pattern generator output termination and protection algorithm.

Controlling the Output Levels

Explains how to set output levels using parameters like Vhigh, Vlow, Vamptl, Voffs.

Allowing the M8020A/M8030A/M8040A to Settle

Explains how instruments react to pattern changes and the time needed to settle.

3 Programming Examples

Initializing the Connection

Shows how to initialize the connection to the instrument using the ResourceName.

4 SCPI Command Language

SCPI Command Language - Introduction

Explains SCPI as the universal programming language for instrument control.

Important Points about SCPI

Covers key areas for SCPI usage: Instrument Model, Command Syntax, and Sending Commands.

Sending Commands to the M8020A/M8030A/M8040A

Details how to send commands, including an overview and using VISA.

5 SCPI Command Reference

Subsystems

Divides SCPI commands into subsystems reflecting instrument functionality.

PLUGin Subsystem

Describes the plugin interface for integrating C# assemblies.

Related product manuals