Keysight M8070A Programming Guide 211
SCPI Command Reference 5
:OUTPut:DATA:F2Jitter:STATe[?]
Syntax :OUTPut:DATA:F2Jitter:STATe ‘identifier’, <ON|OFF|1|0>
:OUTPut:DATA:F2JitterSTATe? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataOut1', 'M*.DataOut2', or 'M*.DataOut'
Return Range ON|OFF|1|0
Description This command enables/disables f/2 jitter at the output. Enabling this
command, causes stop of full clock group (all pattern generator channels
and CLK/TRIG).
This query returns the present setting.
This SCPI is applicable for M8045A.
Example :OUTP:DATA:F2J:STAT ‘M1.DataOut’, ON