216 Keysight M8070A Programming Guide
5 SCPI Command Reference
register. The :PRECursor and :POSTCursor commands change the tap
values only in that preselected preset register. There are 31 preset
registers available addressed by a register index of 0 up to 30.
This command should not be sent in one PM (SCPI - Program Message)
with the following de-emphasis commands:
•:PRESet
• :PRECursor
• :POSTcursor
In other words, the preset register should be selected before sending the
values with the above de-emphasis commands.
Supported preset values for M8041A and M8051A are 0 - 30 and for
M8061A and M8062A, 0 is the only preset value.
This SCPI is applicable for M8041A, M8051A, M8061A and M8062A.
Example :OUTP:DEEM:SEL:PRES ‘M1.DataOut2’, 4