Keysight M8070A Programming Guide 231
SCPI Command Reference 5
:OUTPut:EMBedding:SPARameter[:STATe][?]
Syntax :OUTPut:EMBedding:SPARameter[:STATe] ‘identifier’, <OFF | ON | 0 | 1>
:OUTPut:EMBedding:SPARameter[:STATe]? ‘identifier’
Input
Parameters
‘identifier’:'M*.DataOut1’,‘M*.DataOut2’,‘M*.DataOut3’,or ‘M*.DataOut4’
OFF | ON | 0 | 1
Return Range 0|1
Description This command Switches on/off the compensation via S-parameter file on
the specified output.
The query returns the present setting.
This SCPI is applicable for M8195A and M8196A.
Example :OUTP:EMB:SPAR:STAT ‘M1.DataOut1’, ON
:OUTPut:EMBedding:SPARameter:PROFile[?]
Syntax :OUTPut:EMBedding:SPARameter:PROFile ‘identifier’, <”Filename”>
:OUTPut:EMBedding:SPARameter:PROFile? ‘identifier’
Input
Parameters
‘identifier’:'M*.DataOut1’,‘M*.DataOut2’,‘M*.DataOut3’,or ‘M*.DataOut4’
Description This command selects the S-Parameter profile. S-parameter is defined by
frequency/real/imaginary tuples for 1 to 9 port components.
The query returns the present setting.
This SCPI is applicable for M8195A and M8196A.
Example OUTP:EMB:SPAR:PROF ‘M1.DataOut1’, ”factory/huber_suhner_cable.s2p”
:OUTPut:EMBedding:SPARameter:IPORt[?]
Syntax :OUTPut:EMBedding:SPARameter:IPORt ‘identifier’, <NRf>
:OUTPut:EMBedding:SPARameter:IPORt? ‘identifier’
Input
Parameters
‘identifier’:'M*.DataOut1’,‘M*.DataOut2’,‘M*.DataOut3’,or ‘M*.DataOut4’
Description This command selects the input port in the S-Parameter profile.
The query returns the present setting.