250 Keysight M8070A Programming Guide
5 SCPI Command Reference
:INPut:CDR:FIRSt:TDENsity[?]
Syntax :INPut:CDR:FIRSt:TDENsity ‘identifier’, <NRf>
:INPut:CDR:FIRSt:TDENsity? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<NRf>: Set first order transition density.
Return Range 25% to 100%
Description This command is used to set the first order transition density of the
incoming signal from 25% to 100%. The default is 50%.
This SCPI is applicable for M8041A and M8051A.
Example :INP:CDR:FIRS:TDEN ‘M1.DataIn1’, 25
:INPut:CDR:FIRSt:LBANdwidth[?]
Syntax :INPut:CDR:FIRSt:LBANdwidth ‘identifier’, <NRf>
:INPut:CDR:FIRSt:LBANdwidth? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1' or 'M*.DataIn2'
<NRf>: Set first order loop bandwidth of the CDR.
Return Range 102 kHz to 20 MHz
Description This command is used to set the first order loop bandwidth of the CDR.
The default is 2 MHz. Acceptable units are Hz, kHz, MHz and exponents
(for example, 4E6 is the same as 4 MHz).
This SCPI is applicable for M8041A and M8051A.
Example :INP:CDR:FIRS:LBAN ‘M1.DataIn1’, 4MHz

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals