EasyManua.ls Logo

Keysight Technologies J-BERT M8020A

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 249
SCPI Command Reference 5
:INPut:CDR:RELOck
Syntax :INPut:CDR:RELOck ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn'
Description This command execute a full CDR relock.
This SCPI is applicable for M8062A.
Example :INPut:CDR:RELOck ‘M2.DataIn’
:INPut:CDR:HIGHTD[?]
Syntax :INPut:CDR:HIGHTD ‘identifier’, <ON|OFF|1|0>
:INPut:CDR:HIGHTD? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn'
<ON|OFF|1|0>: Set the High Transition Density state to ON or OFF.
Return Range 0|1
Description This command sets the High Transition Density state to ON or OFF.
This SCPI is applicable for M8062A.
Example :INPut:CDR:HIGHTD ‘M2.DataIn’, 1
:INPut:CDR:OPTimize
Syntax :INPut:CDR:OPTimize ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn'
Description This command invokes the M8062A CDR optimization at the current data
rate and optionally applied jitter.
This SCPI is applicable for M8062A.
Example :INPut:CDR:OPTimize ‘M2.DataIn’