EasyManua.ls Logo

Keysight Technologies J-BERT M8020A - Page 248

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
248 Keysight M8070A Programming Guide
5 SCPI Command Reference
:INPut:CDR[:STATe][?]
Syntax :INPut:CDR[:STATe] ‘identifier’, <ON|OFF|1|0>
:INPut:CDR[:STATe]? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn1', 'M*.DataIn2' or 'M*DataIn'
<ON|OFF|1|0>: Enable/disable the CDR.
Return Range 1|0
Description This command enables or disables CDR.
This SCPI is applicable for M8041A, M8051A and M8046A.
Example :INP:CDR ‘M1.DataIn1’, ON
:INPut:CDR:AUTO[?]
Syntax :INPut:CDR:AUTO ‘identifier’, <ON|OFF|1|0>
:INPut:CDR:AUTO? ‘identifier’
Input
Parameters
‘identifier’: 'M*.DataIn'
<ON|OFF|1|0>: Enable/disable the auto relock.
Return Range 1|0
Description This command enables or disables the auto relock.
This SCPI is applicable for M8062A.
Example The following SCPI command sets the ‘auto relock’ to 1.
:INPut:CDR:AUTO ‘M2.DataIn1’, 1
The following SCPI command queries the ‘auto relock’ state.
:INPut:CDR:AUTO? ‘M2.DataIn1’
1

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Related product manuals