Keysight M8070A Programming Guide 291
SCPI Command Reference 5
:PLUGin:ERATio:ACQuisition:TCLevel[?]
Syntax :PLUGin:ERATio:ACQuisition:TCLevel ‘identifier’, <NRf>
:PLUGin:ERATio:ACQuisition:TCLevel? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set target confidence level.
Return Range 0.1% to 99.9%
Description This command sets the target confidence level when accumulation end is
set to pass/fail (in percent) for the addressed measurement name
identifier. The query returns the current target confidence level.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:ERAT:ACQ:TCL ‘MyMeasurement’, 95
:PLUGin:ERATio:ACQuisition:TERatio[?]
Syntax :PLUGin:ERATio:ACQuisition:TERatio ‘identifier’, <NRf>
:PLUGin:ERATio:ACQuisition:TERatio? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set target error ratio.
Return Range 1E-18 to 1E-3
Description This command sets the target error ratio for the addressed measurement
name identifier. The query returns the current target error ratio value.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:ERAT:ACQ:TER ‘MyMeasurement’, 1e-09

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals