Keysight M8070A Programming Guide 293
SCPI Command Reference 5
:PLUGin:ERATio:SHOW:RVMode[?]
Syntax :PLUGin:ERATio:SHOW:RVMode ‘identifier’, <DET|SUMM>
:PLUGin:ERATio:SHOW:RVMode? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<DET|SUMM>: Select measurement results view mode.
Return Range DET|SUMM
Description This command sets the measurement results view mode to detailed or
summary when viewing the results using the
:PLUGin:ERATio:FETCh:DATA? command. This mode can be changed
during a measurement or after a measurement has completed. The default
is summary mode.
DET (detailed) sets the mode to return the error ratio, compared bits,
errored bits, errored zero ratio, compared zeroes, errored zeroes, errored
one ratio, compared ones, errored ones, etc.
SUMM (summary) sets the mode to return the error ratio, compared bits,
errored bits, etc.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:ERAT:SHOW:RVM ‘MyMeasurement’, DET

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals