344 Keysight M8070A Programming Guide
5 SCPI Command Reference
:PLUGin:OLEVel:ACQuisition:HILevel[?]
Syntax :PLUGin:OLEVel:ACQuisition:HILevel[?] 'Identifier', <NRf>
Input
Parameters
Identifier: 'MyMeasurement'
Return Range -5 V to 5 V
Description This command allows the user to specify the sample threshold high level.
The query returns current value of sample threshold high Level.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:OLEV:ACQ:HIL 'MyMeasurement', 0.5
:PLUGin:OLEVel:ACQuisition:LOWLevel[?]
Syntax :PLUGin:OLEVel:ACQuisition:LOWLevel[?] 'Identifier', <NRf>
Input
Parameters
Identifier: 'MyMeasurement'
Return Range -5 V to 5 V
Description This command allows the user to specify the Sample Threshold Low Level.
The query returns current value of sample threshold low level.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:OLEV:ACQ:LOWL 'MyMeasurement', -0.5