Keysight M8070A Programming Guide 347
SCPI Command Reference 5
:PLUGin:OLEVel:EVALuation:MBQFactor[?]
Syntax :PLUGin:OLEVel:EVALuation:MBQFactor[?] 'Identifier', <NRf>
Input
Parameters
Identifier: 'MyMeasurement'
Range 1e-15…1.0
Description This command allows the user to specify the Minimum BER for Q-factor
caluclations. Please note that in order to be able to do Q-Factor
calculations the specified BER Threshold value must be greater than the
value specified by this command.
The query returns the minimum value specified to allow the Q-Factor
calculations.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:OLEV:EVAL:MBQF 'MyMeasurement', 1E-15