346 Keysight M8070A Programming Guide
5 SCPI Command Reference
:PLUGin:OLEVel:EVALuation Subnode
This subnode has the following SCPI structure:
This subnode has the following commands:
Table 69
:PLUGin:OLEVel:EVALuation:BERThresh[?]
Syntax :PLUGin:OLEVel:EVALuation:BERThresh[?] 'Identifier', <NRf>
Input
Parameters
Identifier: 'MyMeasurement'
Return Range 1E-15 to 1E-1
Description This command allows the user to specify the BER Threshold value against
which the measurement parameters are calculated. The query returns
current value of the BER threshold.
This SCPI is applicable for M8041A, M8051A and M8062A.
Example :PLUG:OLEV:EVAL:BERT 'MyMeasurement', 1E-10
Name Description under
:BERThresh[?] :PLUGin:OLEVel:EVALuation:BERThresh[?]
on page 346
:MBQFactor[?] :PLUGin:OLEVel:EVALuation:MBQFactor[?]
on page 347
:EVALuation
:BERThresh
:MBQFactor