EasyManua.ls Logo

Keysight Technologies J-BERT M8020A - Page 373

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 373
SCPI Command Reference 5
:PLUGin:JTOLerance:BSETup:TBERatio[?]
Syntax :PLUGin:JTOLerance:BSETup:TBERatio ‘identifier’, <NRf>
:PLUGin:JTOLerance:BSETup:TBERatio? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the target BER.
Return Range 1E-15 to 1E-1
Description This command is used to set the target BER of the measurement specified
by the measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:BSET:TBER ‘MyMeasurement’,1E-15
:PLUGin:JTOLerance:BSETup:CLEVel[?]
Syntax :PLUGin:JTOLerance:BSETup:CLEVel ‘identifier’, <NRf>
:PLUGin:JTOLerance:BSETup:CLEVel? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the percent confidence interval.
Return Range 0.1 to 99.9
Description This command is used to set the percent confidence interval of the
measurement specified by the measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:BSET:CLEV ‘MyMeasurement’, 95%

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Related product manuals