374 Keysight M8070A Programming Guide
5 SCPI Command Reference
:PLUGin:JTOLerance:BSETup:FRTime[?]
Syntax :PLUGin:JTOLerance:BSETup:FRTime ‘identifier’, <NRf>
:PLUGin:JTOLerance:BSETup:FRTime? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the frequency relax time.
Return Range 0 ms to 60 s
Description This command is used to set the frequency relax time after a change in
jitter modulation frequency for the measurement specified by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:BSET:FRT ‘MyMeasurement’,30s
:PLUGin:JTOLerance:BSETup:ARTime[?]
Syntax :PLUGin:JTOLerance:BSETup:ARTime ‘identifier’, <NRf>
:PLUGin:JTOLerance:BSETup:ARTime? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the amplitude relax time.
Return Range 0 ms to 60 s
Description This command is used to set the amplitude relax time after a change in
jitter modulation amplitude for the measurement specified by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:BSET:ART ‘MyMeasurement’, 30s

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals