374 Keysight M8070A Programming Guide
5 SCPI Command Reference
:PLUGin:JTOLerance:BSETup:FRTime[?]
Syntax :PLUGin:JTOLerance:BSETup:FRTime ‘identifier’, <NRf>
:PLUGin:JTOLerance:BSETup:FRTime? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the frequency relax time.
Return Range 0 ms to 60 s
Description This command is used to set the frequency relax time after a change in
jitter modulation frequency for the measurement specified by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:BSET:FRT ‘MyMeasurement’,30s
:PLUGin:JTOLerance:BSETup:ARTime[?]
Syntax :PLUGin:JTOLerance:BSETup:ARTime ‘identifier’, <NRf>
:PLUGin:JTOLerance:BSETup:ARTime? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the amplitude relax time.
Return Range 0 ms to 60 s
Description This command is used to set the amplitude relax time after a change in
jitter modulation amplitude for the measurement specified by the
measurement name identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:BSET:ART ‘MyMeasurement’, 30s