Keysight M8070A Programming Guide 375
SCPI Command Reference 5
:PLUGin:JTOLerance:MSETup:TEMPlate:FILE[?]
Syntax :PLUGin:JTOLerance:MSETup:TEMPlate:FILE ‘identifier’, <filepath-string>
:PLUGin:JTOLerance:MSETup:TEMPlate:FILE? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<filepath-string>: Specify path/filename of template to load.
Return Range Path/filename of current measurement template.
Description This command is used to specify the path/filename of the measurement
template to load for the measurement specified by the measurement name
identifier.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUGin:JTOLerance:MSETup:TEMPlate:FILE
"MyMeasurement","Current/default.jtt"
:PLUGin:JTOLerance:MSETup:FREQuency:STARt[?]
Syntax :PLUGin:JTOLerance:MSETup:FREQuency:STARt ‘identifier’, <NRf>
:PLUGin:JTOLerance:MSETup:FREQuency:STARt? ‘identifier’
Input
Parameters
‘identifier’: Specify the measurement name.
<NRf>: Set the start frequency.
Return Range 100 Hz to 500 MHz
Description This command is used to set the start (minimum) modulation frequency for
the measurement specified by the measurement name identifier. The
default is 100 kHz.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A and
M8046A.
Example :PLUG:JTOL:MSET:FREQ:STAR ‘MyMeasurement’, 10kHz