Keysight M8070A Programming Guide 417
SCPI Command Reference 5
:PLUGin:EDIagram:SHOW:WAVeform:THReshold[?]
Syntax :PLUGin:EDIagram:SHOW:WAVeform:THReshold 'Identifier', <NR3>
:PLUGin:EDIagram:SHOW:WAVeform:THReshold? 'Identifier'
Input
Parameters
Identifier: 'MyMeasurement'
Return Range 1E-6 to 1E-1
Description This command sets the waveform BER threshold value.
This query returns the current setting.
This SCPI is applicable for M8041A, M8051A and M8062A
Example :PLUG:EDI:SHOW:WAV:THR 'MyMeasurement', 1E-3