EasyManua.ls Logo

Keysight Technologies J-BERT M8020A - Page 435

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Keysight M8070A Programming Guide 435
SCPI Command Reference 5
:PLUGin:PSWEep:SHOW:SMPoints[?]
Syntax :PLUGin:PSWEep:SHOW:SMPoints 'Identifier', <0|1|ON|OFF>
:PLUGin:PSWEep:SHOW:SMPoints? 'Identifier'
Input
Parameters
Identifier: Specify the plugin name. For e.g. 'ParameterSweep1'
<OFF|ON|0|1>: Disable/enable viewing the measurement points.
Return Range 0|1
Description This command disables/enables viewing of the measurement points in the
GUI and associates it with a specific measurement name identifier. A 0 or
OFF value disables viewing and a 1 or ON value enables viewing of the
measurement points.
This query returns the current setting.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example :PLUG:PSWE:SHOW:SMP 'ParameterSweep1', 1
:PLUGin:PSWEep:SHOW:VSCale[?]
Syntax :PLUGin:PSWEep:SHOW:VSCale 'Identifier', <LOGarithmic | LINear >
:PLUGin:PSWEep:SHOW:VSCale? 'Identifier',
Input
Parameters
Identifier: Specify the plugin name. For e.g. 'ParameterSweep1'
Return Range LOG | LIN
Description This command specifies the vertical BER axis scale and associates it with a
specific measurement name identifier. If LOGarithmic is specified then the
BER values are in terms of a logarithmic scale. If LIN is specified then the
BER values are in terms of a linear scale.
This query returns the current setting.
This SCPI is applicable for M8041A, M8051A, M8061A, M8062A, M8045A
and M8046A.
Example :PLUG:PSWE:SHOW:VSC 'ParameterSweep1', LOG

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Related product manuals