470 Keysight M8070A Programming Guide
5 SCPI Command Reference
:SYSTem:RESet[:ALL]
Syntax :SYSTem:RESet[:ALL] [HARD | SOFT]
Input
Parameters
[HARD | SOFT]
Description This command resets the instrument.
If SOFT is passed the instrument will load its default values and defined
'Groups' in the GUI will be untouched.
If HARD is passed the instrument will load the factory setting and all
'Group' definitions will be cleared.
Independent of passing HARD or SOFT all plug-ins (measurements) are
closed.
Example :SYSTem:RESet:ALL SOFT
:SYSTem:RESet:INSTrument
Syntax :SYSTem:RESet:INSTrument [HARD | SOFT]
Input
Parameters
[HARD | SOFT]
Description This command resets the instrument.
If SOFT is passed the instrument will load its default values and defined
'Groups' in the GUI will be untouched.
If HARD is passed the instrument will load the factory setting and all
'Group' definitions will be cleared.
Example :SYSTem:RESet:INSTrument SOFT

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals