Keysight M8070A Programming Guide 473
SCPI Command Reference 5
:DATA:PATTern:USE[?]
Syntax :DATA:PATTern:USE <pattern-name>,<num-symbols>[,<symbol-coding>
[,<mask-used>[,<squelch-used>,<error-used>]]]
:DATA:PATTern:USE? <pattern-name>
Input
Parameters
<pattern-name>: Name of the pattern. Name consists of root node
(“factory:”, “shared:”, or “current:”) plus folder names separated by ‘/’ plus
pattern name.
<num-symbols>: Length of the pattern in number of symbols.
<symbol- coding>: BIT|B8B10|B128B130|B128B132 for bit coding, 8b10b
coding, 128/130 coding or 128/132 coding. Default is BIT.
<mask- used>: 0|1|OFF|ON to specify if mask attribute is used. Default is
OFF. Not using mask is mostly useful for bit coding to save space.
<squelch-used>: 0|1|OFF|ON to specify if squelch attribute is used. Default
is OFF. Not using squelch is mostly useful for bit coding to save space.
<error-used>: 0|1|OFF|ON to specify if error attribute is used. Default is
OFF. Error Attribute is used for captured patterns. When <error-used> is
ON, <mask-used> and <squelch-used> must be OFF.
Return Range <num-symbols>, <symbol-coding>, <mask-used>,
<squelch-used>,<error-used>
Description This command creates a new pattern. If a pattern exists with the same
name, it is overwritten.
Example :DATA:PATT:USE "current:demo",20
Example :DATA:PATT:USE "current:demo8b10b",5,b8b10
Example :DATA:PATT:USE "current:demo128b130b",5,b128b130
Example :DATA:PATT:USE "current:demo128b132b",5,b128b132
Example :DATA:PATT:USE? "current:demo"
5,BIT,0,0,0
Example :DATA:PATT:USE? "factory:SATA/LTDP_short_b8b10"
512,B8B10,1,1,0