484 Keysight M8070A Programming Guide
5 SCPI Command Reference
:DATA:SEQuence:DELete
Syntax :DATA:SEQuence:DELete <sequence-name>
Input
Parameters
<sequence-name>: Specify sequence name.
Description This command deletes the specified sequence.
Example :DATA:SEQ:DEL ‘MySequence’
:DATA:SEQuence:DALL
Syntax :DATA:SEQuence:DALL
Description This command deletes all sequences. This is a useful function if you do not
want to use the reset sequence (“Generator”, “Analyzer”).
Example :DATA:SEQ:DALL
:DATA:SEQuence:BREak
Syntax :DATA:SEQuence:BREak ‘identifier’
Input
Parameters
<identifier>: ‘M*.System’
Description This command sends a break signal addressed by the identifier. This will
leave a loop and execute the next block or execute an arbitrary jump to
another block.
Example :DATA:SEQ:BRE ‘M1.System’
:DATA:SEQuence:RESTart
Syntax :DATA:SEQuence:RESTart <sequence-name>
Input
Parameters
<sequence-name>: Specify sequence name.
Description This command restarts the specified sequence (jumps to the first block).
For example, this command is useful for ED sequences. Usually there is a
synch block first and then a loop to compare the synchronized pattern.
Performing a restart there causes a synchronization of the sequence.

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish

Related product manuals