EasyManuals Logo

Keysight Technologies J-BERT M8020A User Manual

Keysight Technologies J-BERT M8020A
502 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #485 background imageLoading...
Page #485 background image
Keysight M8070A Programming Guide 485
SCPI Command Reference 5
Example :DATA:SEQ:REST ‘EdSequence’
:DATA:SEQuence:RESTart:IDENtifier
Syntax :DATA:SEQuence:RESTart:IDENtifier <identifier>
Input
Parameters
<identifier>: 'M*.DataIn1', 'M*.DataIn2', 'M*.DataIn', 'M*.DataOut',
'M*.DataOut1' and 'M*.DataOut2'
Description This command restarts the specified sequence (jumps to the first block).
For example, this command is useful for ED sequences. Usually there is a
synch block first and then a loop to compare the synchronized pattern.
Performing a restart there causes a synchronization of the sequence.
Example :DATA:SEQ:REST:IDEN ‘M1.DataIn1’
:DATA:SEQuence:BLOCk?
Syntax :DATA:SEQuence:BLOCk? <sequence-name>
Input
Parameters
<sequence-name>: Specify sequence name.
Description This command returns the block of sequences currently executed.
Example :DATA:SEQuence:BLOCk? ‘Generator’
:DATA:SEQuence:BLOCk:IDENtifier?
Syntax :DATA:SEQuence:BLOCk:IDENtifier? <identifier>
Input
Parameters
<identifier>: 'M*.DataIn1', 'M*.DataIn2', 'M*.DataIn', 'M*.DataOut',
'M*.DataOut1' and 'M*.DataOut2'
Description This command returns the block of sequences currently executed for a
specific location.
Example :DATA:SEQ:BLOC:IDEN? ‘M1.DataIn1’

Table of Contents

Other manuals for Keysight Technologies J-BERT M8020A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keysight Technologies J-BERT M8020A and is the answer not in the manual?

Keysight Technologies J-BERT M8020A Specifications

General IconGeneral
BrandKeysight Technologies
ModelJ-BERT M8020A
CategoryControl Unit
LanguageEnglish