492 Keysight M8070A Programming Guide
5 SCPI Command Reference
Range Symbol 1 Level: 1% to 65%
Symbol 2 Level: 34% to 99%
Description The command controls the PAM-4 level mappings. It can be used to adjust
the actual output level that is being generated by a data output for a
specific PAM-4 symbol.
The levels of PAM-4 symbol 0 and 3 cannot be adjusted and are always
kept at the value of 0% and 100% of the actual output voltage swing.
It is best practice to configure symbol level 1 and 2 in a combined remote
command transaction (see example below) to minimize hardware
reconfiguration duration. This is especially important when the location is
addressing a data output that is located on an arbitrary waveform
generator module (e.g. M8195A).
This query returns the present state.
This SCPI is applicable for M8195A, M8196A and M8045A.
Example :DATA:LIN:PAM4:LEV1 ‘M1.DataOut1’, 35pct;LEVel2 ‘M1.DataOut1’, 70pct