l The maximum repeat count for CD ARBs is 256. See [SOURce:]ARB:COUNt.
[SOURce:]ARB:CURRent:CDWell:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:CDWell:POINts? (@<chanlist>)
Returns the number of current or voltage points in a constant-dwell Arb in the form +n.nnnnnnE+nn.
Parameter Typical Return
(none) <number of points>
Program a constant-dwell time of 0.2 seconds:
ARB:CURR:CDW:DWEL 0.2,(@1)
ARB:VOLT:CDW:DWEL 0.2,(@1)