Parameter Typical Return
Current: 0 - maximum
Voltage: 0 - maximum
The maximum value varies, depending on the currentor voltage
rating of the power module.
*RST 1 pointsetto the minimum value.
<value> [,<value>]
Step 0 - 99, *RST 0 (none)
Program the current points: ARB:CURR:UDEF 5,4,3,2,1,(@1)
Program the voltage points: ARB:VOLT:UDEF 20,21,22,23,24,(@1)
Program the points for step 0: ARB:SEQ:STEP:CURR:UDEF5,4,3,2,1,0,(@1)
l At the end of the user-defined Arb, the output state of the unit depends upon the
[SOURce;]ARB:TERMinate:LAST program settings.
l The order in which the values are entered determines the sequence of execution.
l To create a valid user-defined Arb, the current level, voltage level, BOST, and dwell lists must
either all be the same length, or have a length of 1, which is interpolated as having the same
length as the maximum length list.
[SOURce:]ARB:CURRent:UDEFined:LEVel:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:LEVel:POINts? (@<chanlist>)
Returns the number of current or voltage points in the user-defined Arb, not the point values.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:LEVel:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:LEVel:POINts? <step#>, (@<chanlist>)
Returns the number of current or voltage points in the user-defined Arb for Arb sequence steps.
Parameter Typical Return
(none) <number of points>
Step 0 - 99, *RST 0 (none)
Returns the number of currentpoints: ARB:CURR:UDEF:LEV:POIN? (@1)
Returns the number of voltage points: ARB:VOLT:UDEF:LEV:POIN? (@1)
Returns the number of points for step 0: ARB:SEQ:STEP:CURR:UDEF:LEV:POIN? 0,(@1)
7 SCPI Programming Reference
262 Keysight N6705C Operating and Service Guide