l At the end of the user-defined Arb, the output state of the unit depends upon the
[SOURce;]ARB:TERMinate:LAST program settings.
l The order in which the values are entered determines the sequence of execution.
l To create a valid user-defined Arb, the current level, voltage level, BOST, and dwell lists must
either all be the same length, or have a length of 1, which is interpolated as having the same
length as the maximum length list.
[SOURce:]ARB:CURRent:UDEFined:DWELl:POINts? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:DWELl:POINts? (@<chanlist>)
Returns the number of dwell points in the user-defined Arb, not the point values.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:DWELl:POINts? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:DWELl:POINts? <step#>, (@<chanlist>)
Returns the number of dwell points in the user-defined Arb for Arb sequence steps.
Parameter Typical Return
(none) <number of points>
Step 0 - 99, *RST 0 (none)
Returns the number of currentdwell points: ARB:CURR:UDEF:DWEL:POIN? (@1)
Returns the number of currentdwell points: ARB:VOLT:UDEF:DWEL:POIN? (@1)
Returns the number of dwell points for step 0: ARB:SEQ:STEP:CURR:UDEF:DWEL:POIN? 0,(@1)
[SOURce:]ARB:CURRent:UDEFined:LEVel <value>{<,value>},(@<chanlist>)
[SOURce:]ARB:CURRent:UDEFined:LEVel? (@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:LEVel <value>{<,value>},(@<chanlist>)
[SOURce:]ARB:VOLTage:UDEFined:LEVel? (@<chanlist>)
Specifies the level of each current or voltage point in a user-defined Arb. A comma-delimited list of up
to 511 points may be programmed.
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:LEVel <value>{<,value>}, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:CURRent:UDEFined:LEVel? <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:LEVel <value>{<,value>}, <step#>, (@<chanlist>)
[SOURce:]ARB:SEQuence:STEP:VOLTage:UDEFined:LEVel? <step#>, (@<chanlist>)
Specifies the level of each current or voltage point in a user-defined Arb for Arb sequence steps. The
last value in the sequence is the step number.